Lap Shear Strength of Adhesively Bonded Metal Specimens (ASTM D1002)

The strength of adhesive bonds depends greatly on the conditions under which the bonding process is carried out. Therefore, variations in this process lead to variations in the accuracy of the results acquired from strength tests of these bonds. It is the responsibility of the manufacturer to prepare samples in accordance with the ASTM standard.

The ASTM standard specifies that “the testing machine shall conform to the requirements of Practices E 4.” For this test, we used the 5569 electromechanical test frame, with a 50 kN load cell and 30 kN capacity mechanical wedge action grips with coarse serrated faces. The standard further specifies that “the long axis of the test specimen coincide with the direction of applied pull through the center line of the grip assembly.” In order to ensure that the load be applied in exactly the same direction as the center line, we inserted spacers into the grips. Ideally the spacers would be bonded to, or machined into, the specimen.


À propos de cette solution

Normes associées ASTM D1002 | ISO 4587 | JIS K6850
Type d’éprouvette : Bande
Matériaux : Métaux
Type d’essai : Pelage, Arrachement et Friction
Secteur d'activité : Sans objet

You are viewing a static image in place of an informative flash animation. Please download Adobe® Flash Player and refresh this page to view the animation.
Intéressé par cette solution ?
Contactez-nous

Documents associés

ASTM D1002 Lap Shear   ASTM D1002 Lap Shear
Série 5900 - La confiance en vos résultats   Série 5900 - La confiance en vos résultats

Nos labos

Les laboratoires d'application Instron servent à la recherche et au développement, tout comme aux essais sur les éprouvettes de nos clients et à la définition de l'équipement adapté.
Prenez contact avec nos spécialistes d'application pour définir la solution qui vous convient le mieux !